gas-lab Q1
Flameless gas quality analyser for natural gas which works on base of a novel infrared absorption technique and heat conductivity. The set up consists of two parts. A heavy duty housing accommodates the sensors and the primary measurement electronics and can optionally be installed in a hazardous environment.
| Technical Data | Downloads | ||||
gas-lab Q1: gas quality analyser
Subject to change without prior notice
| Primary measurements and ranges |
Gross calorific value Hs 8.4 -13.1 kWh/m³ (30.4 - 47.2 MJ/m³) Standard density ρ 0.711 - 0.970 kg/m³ CO2-concentration xCO2 0 - 5 mol%, optional 0 - 20 mol% (Reference state: combustion 25°C, volume 0°C, othjer reference states optional) |
| Measurement uncertainty |
Hs: < 0.4%, ρ: < 0.8%, xCO2:< 0.2 mol% Hs: < 0.2%, ρ: < 0.4% optional with individual gas calibration |
| Derived measurements |
Wobbe index, net CV, rel. density, sat. gross/net CV, methane number and calculated model gas composition (CH4...C8H18, N2, CO2) |
| Range of appropiate gases |
Generic natural gases CH4: 75- 100 mol% CO2: 0- 5 mol%, optional 0 - 20 mol% C2+ 0- 15 mol% O2: 0 - 3 mol% N2 0- 20 mol% others < 0.1 mol% |
| Calibration gas | Methane (purity > 3.5) with automatic calibrationcycle |
| Gas management |
Integrated 3-channel double block an bleed device for the process gas, calibration gas and optional test gas (remote verification possible) |
| Operational pressure/consumption |
Inlet pressure 200 mbar to 5 bar (basic configuration), 30 l/h (without bypass) Inlet pressure 200 mbar to 100 bar with optional high pressure reductin unit |
| Dynamics | Analysis rate: c. 1/s, reaction time < 5 - 10 s, t90: c. 20 - 60 s (bypass setting) |
| Power supply | 24 V DC, 30 W |
| Enviroment |
Sensor unit: IP 54, 5 - 40 °C, with additional housing: -20 - 55 °C Controller unit: IP 20, 0 - 40 °C |
| Interfaces |
3 digital outputs, 4 analogue outputs, optionally expandable I/O boards, serial interface forparameter setting and local readout, DSfG interface, optional Modbus interface, optional modem for remote controlan data retrieval |
| Data logger |
Integrated logging of Hs, ρ, xCO2 and derived measurement as mean values (e.g. on hourly basis) or current values, Integrated logging of measurement system status and external events, Data retrieval locally vie serial I/O or remotely via optional modem |
| Approvals | ATEX, EMC, PTB (National institute for metrology in Germany) |